Terminated

Xper Flex Cardio Physiomonitoring System Model # 453564669081 Software Version 1.6.0.0388

FDA Recall Z-3192-2018 · Posted None

Recalling FirmInvivo Corporation
Recall StatusTerminated
Product CodeMWI
Date Initiated2018-08-03
Date Terminated2021-07-15
Product Quantity38

Reason for Recall

Released with an incorrect, older firmware version installed. May result in intermittent loss of ECG monitoring, inaccurate ECG amplitude display and or auxiliary output, and inaccurate heart rate (HR) displayed due to QRS detection fault.

Firm Action

Philips is asking customers to follow the Action to be Taken by Customer/User section of the August 3, 2018, Field Safety Notice: The Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. Philips will replace all affected devices. Users should discontinue use of affected devices immediately. A Philips representative will contact you regarding your affected device.

Root Cause

Software Manufacturing/Software Deployment

Distribution Pattern

Distribution in the Netherlands

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.