Terminated

Varian Medical Systems Inc., Millennium and HD-120 MultiLeaf Collimators (MLC), a component of the Clinac, All Varian Clinac using the collimator cover P/N 1106900-xx, Model Numbers: H14, H18, H27, H29, HCX; part of the Trilogy Radiotherapy Delivery System.

FDA Recall Z-3204-2011 · Posted 2011-09-14

Recall StatusTerminated
Product CodeIYE
Date Initiated2011-03-17
Date Terminated2012-06-20
Product Quantity3791 units

Reason for Recall

An improperly installed collimator cover may detach and fall from the machine, potentially striking the patient or nearby personnel.

Firm Action

On 3/17/2011, Varian Medical Systems sent an Urgent Medical Device Correction Letter to all affected users, with a description of the problem and user corrective action steps.

Root Cause

Component design/selection

Distribution Pattern

Worldwide distribution including USA, Australia, Argentina, Italy, Morocco, South Korea, Belgium, Canada, Netherlands, Denmark, Japan, Hong Kong, Malaysia, Hungary, New Zealand, Sweden, Taiwan, Switzerland, Romania, Brazil, Slovakia, India, Colombia, UK, Ireland, Spain, Singapore, Martinique, Reunion, Venezuela, Iceland, Portugal, Turkey, Israel, Bahrain, Macau, Chile, Saudi Arabia, Latvia, China, Finland, Kazakhstan, Thailand, France, Poland, and Germany.