TRIDENT 0 POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)
FDA Recall Z-3204-2026 · Posted 2026-09-16
Recalling FirmHowmedica Osteonics Corp.
Recall StatusOpen, Classified
Product CodeMEH
Date Initiated2026-07-30
Product Quantity12,156 US
Reason for Recall
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Firm Action
Firm began notifying customers on July 30, 2026, via FedEx 2-day air.
Customers were instructed to quarantine and discontinue use of the affected product. Affected product should be returned to the firm.
Root Cause
Process change control
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.