Open, Classified

CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-7600-24, 21-7609-24 OUS Product Codes: 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24JP 21-7600-24JP 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24

FDA Recall Z-3212-2026 · Posted 2026-09-17

Recalling FirmICU Medical, Inc.
Recall StatusOpen, Classified
Product CodeLHI
Date Initiated2026-08-20
Product Quantity185,688 units

Reason for Recall

Reservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or air embolus.

Firm Action

On August 20, 2026, the firm began notifying affected customers via Urgent Medical Device Correction letters. Customers were informed that certain reservoirs may exhibit fluid leakage at the joint between the connector and the tubing. Customers were instructed to check all inventory locations for affected product and discontinue use. All affected products should be destroyed following their institution's destruction process. If destruction is not immediately possible at your facility, quarantine the products until disposal. Customers should share the notification with all potential users of the reservoirs, and notify any accounts to whom the affected products have been further distributed. The firm will provide replacement products to affected customers upon receipt of a completed Customer Response Form to certify product destruction.

Root Cause

Process control

Distribution Pattern

Worldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN, LA, MI, MO, MS, TN, TX, VA, WI and the countries of Austria, Australia, Belgium, Costa Rica, Germany, Estonia, Spain, Finland, France, Great Britain, Ireland, Italy, Japan, Kuwait, Netherlands, Norway, and Sweden.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →