Sarns 8000 Cardioplegia Monitor, Catalog Number 16414; also known as PROG COMP BD AMPRO 420 CPG 8K, Catalog Number 816373 The Sarns 8000 Cardioplegia Monitor is indicated to measure extracorporeal line pressure and signal when the pressure exceeds the operator-set limit.

FDA Recall Z-3214-2011 · Posted 2011-09-15

Recall StatusTerminated
Product CodeDWB
Date Initiated2011-08-29
Date Terminated2013-03-08
Product Quantity85

Reason for Recall

The Cardioplegia Monitor for the Sarns Modular Perfusion System 8000 may fail to track cardioplegia delivery volume.

Firm Action

Terumo Cardiovascular Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated August 29, 2011 to all affected customers. The letter describes the product, problem, and action to be taken. The letter advises customers to make sure all users are aware of the notice. Customers are instructed to complete and return an attached Customer Response Form. Terumo will replace the software on all monitors of the affected product. Customers with questions or concerns should call Terumo CVS Customer Service 1-800-521-2818.

Root Cause

Component change control

Distribution Pattern

Worldwide Distribution--USA (nationwide) including the states of AL, AR, CA, CO, GA, IL, IN, MS, NC, OH, TN, VA, and WI and countries of Belgium, Columbia, Hong Kong, India, Singapore, and Taiwan.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑