Ultradent Products, Inc. Paper Points, Size 50 PN #1559, 200 per container, sterile. Product Usage: The product is used to assure that all moisture is removed from the root canal before the obturation procedure begins..

FDA Recall Z-3218-2017 · Posted None

Recall StatusTerminated
Product CodeEKN
Date Initiated2017-08-04
Date Terminated2018-01-04
Product Quantity62 units

Reason for Recall

The barcode label was placed underneath the sterile barrier and the product was re-wrapped.

Firm Action

The recalling firm issued letters dated August 4, 2017 via certified mail,2-day delivery mail, or email on August 4, 2017. The letter identified the affected product, problem, and actions to be taken. For questions contact Regulatory Affairs Specialist at Ultradent Products, Inc.

Root Cause

Process change control

Distribution Pattern

Distribution was nationwide. Foreign distribution was made to Canada, Bulgaria, Namibia, New Zealand and Germany. There was no military /government distribution.