FUJINON ED-530XT. This device is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.
FDA Recall Z-3225-2017 · Posted None
Advertisement
Recalling FirmFujifilm Medical Systems U.S.A., Inc.
Recall StatusTerminated
Product CodeFDT
Date Initiated2017-07-21
Date Terminated2018-07-12
Product Quantity362 units
Reason for Recall
An update to the design and labeling was implemented to help reduce patient risk associated with inadequate reprocessing of the device.
The action includes replacement of the forceps elevator mechanism, the O-ring seal, the distal end cap and issuance of a new Operation Manual.
Firm Action
On July 21, 2017, Fujifilm Medical Systems U.S.A., Inc. Endoscopy Division distributed Urgent Medical Device Correction and Removal notices & Field Action Verification Forms to their U.S. Consignees by USPS certified mail. Fujifilm Medical Systems will contact all customers that have the duodenoscope by October 2017, with details for returning the device for replacement parts. This field action allows continued use of the current device until you are contacted by a sales representative for replacement parts as early as October 2017. While the remediation of the scopes will take approximately two weeks to complete, Fujifilm Medical Systems U.S.A. Inc. Endoscopy Division will provide interim duodenoscopes on loan for continuity of care. Once received, the recalling firm will replace the forceps elevator mechanism with O-ring and the distal end cap. An updated manual has been generated to include new information/guidance. Customers are encouraged to complete and return the Field Action Verification Form via email to [email protected] or via fax to (973) 872-4723. Customers with questions can call Larry Picciano, Sr. Director Quality and Regulatory Affairs at (973) 686-2479.
Root Cause
Device Design
Distribution Pattern
Nationwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.