Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/ MDR107002E MDR520094 PENDANT F/ MDR107003E MDR530094 PENDANT, W/ 9V BATTERY BACKUP, F/ 2E BED MDR540094 PENDANT, W/ 9V BATTERY BACKUP, F/ 3E BED

FDA Recall Z-3227-2026 · Posted 2026-10-01

Recall StatusOpen, Classified
Product CodeLLI
Date Initiated2026-08-14
Product Quantity771,127 total

Reason for Recall

Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.

Firm Action

Firm initially notified customers November 26, 2025. On August 14, 2026, the firm sent a follow-up letter which provides additional product SKU - Item numbers for the electrical system safety risks and additional important information related to the previous notices. The notice explained the issue, potential risks, and requested the user ONLY USE MEDLINE SIDE RAILS and MEDLINE ACCESSORIES that are intended for use with Medline Homecare Beds. When using Medline Side Rails and Medline Accessories, follow all applicable instructions for use. Distributors were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 [email protected].

Root Cause

Device Design

Distribution Pattern

Distribution US Nationwide including territories of Puerto Rico and Guam and the countries of Bermuda, Canada, and Panama.