Terminated

Durex Real Feel Polyisoprene Condoms, 10 count and 24 count, SKU Numbers 02340-89456-14 and 02340-89457-00 Product Usage: The Durex Synthetic polyisoprene male condom is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted disease).

FDA Recall Z-3241-2018 · Posted None

Recalling FirmReckitt Benckiser LLC
Recall StatusTerminated
Product CodeMOL
Date Initiated2018-07-23
Date Terminated2020-03-05
Product Quantity487980 condoms

Reason for Recall

The products failed the filed specification of 1.0kPa for Burst Pressure.

Firm Action

On 7/23/18, an Urgent Medical Device Recall notification letter was distributed to customers. The letter instructed customers to do the following: 1. Examine your inventory in your distribution center immediately to determine if you have the specified batch in your warehouse inventory. 2. Segregate and quarantine the specified batch. 3. If you have product of the specified batch, contact our Stericycle representative immediately for a RETURN AUTHORIZATION number. a. Contact Information: Phone: 877-448-5308 Fax: 877-884-9410 4. Contact Stericycle Expert Solutions at 877-448-5308 for assistance with coordination of pickup and removal of product from your facility. If arranging pick up of product, freight will be prepaid by Stericycle. A pre-paid return label has been included for your convenience. Alternatively, you may arrange shipment of product directly to Stericycle at the following address: Attn: Event 4241 2670 Executive Drive, Suite A Indianapolis, IN 46241 Dock Hours are SAM -4:30PM Monday- Friday Please call to schedule a dock appointment 5. Finally, notify Stericycle representative via email [email protected] that you have: a. Checked your inventory in your distribution centers and confirm you have NO product of the specified batch b. You have collected all units in your inventory and product has been shipped back in accordance with the instructions from your Stericycle representative. On 7/26/18, a second Urgent Medical Device Recall notification letter was sent to customers which included two additional batches included in the recall. This letter included the same customer instructions as the 7/23/18 letter.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide.