Terumo 100/120V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801763
FDA Recall Z-3242-2018 · Posted None
Advertisement
Recalling FirmTerumo Cardiovascular Systems Corporation
Recall StatusTerminated
Product CodeDTQ
Date Initiated2018-08-28
Date Terminated2020-04-23
Product Quantity47 units
Reason for Recall
Screws on a Terumo Advanced Perfusion
System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system.
Firm Action
Terumo CVS issued letter on 8/2/818 via Fed'X 2 day air to inform users of the problem for the Power Manager boards with unverifiable or out of specification screws, potential hazard and field correction. A Terumo Field Service Representative will contact users to schedule the field correction activities. Terumo CVS Customer Service: 1.800.521.2818, Monday Friday, 8 a.m. 6 p.m. ET
Root Cause
Process control
Distribution Pattern
Nationwide Foreign: Mexico, BELGIUM, CANADA, India, Indonesia, Singapore, South Korea , Vietnam