Terminated

Visualase Thermal Therapy System, Software: Visualase Software versions 3.1.1 - 3.2, Product #: 9735542

FDA Recall Z-3251-2018 · Posted None

Recall StatusTerminated
Product CodeLLZ
Date Initiated2018-06-01
Date Terminated2022-02-02
Product Quantity161

Reason for Recall

Inaccuracy of MR thermometry during MRI-guided laser ablation procedures using the Medtronic Visualase Thermal Therapy System may result in unaccounted spread of thermal energy to the surrounding tissue.

Firm Action

On 06/01/2018, Urgent Medical Device Correction letters and Consignee Response Forms were mailed to customers via overnight mail service. Customers were provided with additional information intended to supplement the system manual and instructed to call the firm at 1-888-826-5603 if they had any questions.

Root Cause

Under Investigation by firm

Distribution Pattern

U.S.: MN, NY, PA, FL, AZ, MA, VA, CA, CO, AL, WI, TX, MO, DC, OH, NH, GA, SC, NJ, WA, CT, MI, OK, IN, MD, IL, KY, LA, OR, RI, UT, AR, MS, TN, NC Foreign: United Kingdom, Germany, Switzerland, Netherlands, Israel