Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, US; Product Code: 0042-0010-US. 7. Optical, AIC, Impella Connect, Pkgd, US; Product Code: 0042-0040-US.

FDA Recall Z-3251-2026 · Posted 2026-09-21

Recalling FirmAbiomed, Inc.
Recall StatusOpen, Classified
Product CodeOZD
Date Initiated2019-02-19
Product Quantity3,322 units

Reason for Recall

Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in Ventricle alarm in certain AICs.

Firm Action

No customer notifications were sent. The retrospectively reported recall does not currently affect any Automated Impella Controllers (AICs) in the field pending the correction. The hardware update was implemented to AICs in the field through servicing to address potential safety concerns.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide - US Nationwide and the countries of Austria, Belgium, Canada, France, Germany, Hong Kong, Italy, Luxembourg, Poland, Spain, Switzerland, United Kingdom.

Official Source

View FDA Recall →