nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT 8100EP1-ST, CAT 8100EP1-ST-US, CAT 8100EP1-US, CAT 8100EP1R

FDA Recall Z-3253-2026 · Posted 2026-09-22

Recall StatusOpen, Classified
Product CodeMHX
Date Initiated2026-05-06
Product Quantity1,007

Reason for Recall

Complaint trend where "Checkline Error" alarm presents at device setup, rendering the CO2 monitoring function inoperable. May result in procedure delay.

Firm Action

Firm posted its initial notice on May 6, 2026. A follow-up letter was sent on August 31, 2026 with additional instructions. Customers were instructed to do the following: 1. Be aware of the Checkline Error alarm. If it appears on your CO2 section at startup, follow the steps provided in the letter. 2. Verify your backup plan. Confirm that your practice has access to an alternate CO2 monitoring source. 3. Retain the letter and note your unit s serial number for your records. 4. Contact the firm if you have any questions.

Root Cause

Process design

Distribution Pattern

US Nationwide distribution including in the states and territory of AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NH, NJ, NV, NY, OH, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and Puerto Rico.