Terminated

7 F Therapy Ablation Catheter, 4 mm Tip Thermistor, Model (Reorder #) 83726, Description: 1304-7-25-XL-TH. St Jude Medical Company A sterile, single use catheter with one 4mm ablation electrode at the tip and three 2mm diagnostic electrodes. The catheter includes a temperature sensor in the tip electrode for temperature monitoring and the handle is equipped with a steering mechanism to detect the distal tip of the catheter; both uni-directional and bi-directional versions are available. The catheter is available with various distal curve configurations; the curve is indicated on the catheter label.

FDA Recall Z-3257-2011 · Posted 2011-09-16

Recalling FirmSt Jude Medical
Recall StatusTerminated
Product CodeLPB
Date Initiated2011-02-10
Date Terminated2012-04-03
Product Quantity179 units total

Reason for Recall

The recall was initiated because St. Jude Medical has confirmed that the Therapy Ablation Catheter, 4 mm Tip Thermistor Quadripolar, were packaged with the incorrect Instructions for Use (IFU). To ensure accuracy of device labeling, St. Jude Medical is requesting that the incorrect IFU be replaced with the enclose, correct version.

Firm Action

St. Jude Medical sent an "important Medical Device Safety Information" letter dated February 22, 2011 to all affected customers. The letter includes: affected problem, reconciliation form and corrective action. Customers are asked to remove the incorrect Instructions for Use (IFU) and use the IFU provided in the letter, and to fill out and return the reconciliation form. For questions on this recall please contact the firm at (512) 329-1054 or (651) 756-2869.

Root Cause

Process control

Distribution Pattern

Nationwide Distribution including the states of TN, MS, IN, MO, FL, WI, CA, and MT.