Medline Convenience Kits: 1) ANTERIOR SPINE PACK, Catalog Number: DYNJT10197; 2) CRANIOTOMY #IMF 56646-LF, Catalog Number: DYNJVB91001; 3) CRANIOTOMY PACK-LF, Catalog Number: DYNJ0369348O; 4) SLUH CRANIOTOMY PACK-LF, Catalog Number: DYNJ0369348P; 5) LAMINECTOMY PACK-LF, Catalog Number: DYNJ04342M; 6) LAMINECTOMY PACK-LF, Catalog Number: DYNJ43218G; 7) CRANIOTOMY PACK, Catalog Number: DYNJ46024F; 8) LAMINECTOMY PACK-LF, Catalog Number: DYNJ46070C; 9) CRANI PACK, Catalog Number: DYNJ46672B; 10) SPINE TRAY II, Catalog Number: DYNJ52961I; 11) SPINE PACK, Catalog Number: DYNJ53518C; 12) SPINE PACK, Catalog Number: DYNJ54479C; 13) CRANI ADJUNCT PACK, Catalog Number: DYNJ57815; 14) WC SPINE PACK-LF, Catalog Number: DYNJ57971C; 15) PEDI CRANI MAJOR, Catalog Number: DYNJ59338C; 16) LAMINECTOMY PACK-LF, Catalog Number: DYNJ66742C; 17) SPINAL PACK-LF, Catalog Number: DYNJ67543B; 18) LAMINECTOMY PACK, Catalog Number: DYNJ69214A; 19) EPCH LAMINECTOMY I PACK, Catalog Number: DYNJ83098; 20) SPINE, Catalog Number: DYNJ900269I; 21) PACK,CRANIOTOMY, Catalog Number: DYNJ906893C; 22) PACK,CRANIOTOMY, Catalog Number: DYNJ906893D; 23) PACK,CRANIOTOMY, Catalog Number: DYNJ906893F; 24) UNIVERSITY CRANIOTOMY, Catalog Number: DYNJ907192B; 25) UNIVERSITY CRANIOTOMY, Catalog Number: DYNJ907192C; 26) DR MOSHEL CRANI, Catalog Number: DYNJ907920C; 27) SPINE-LF, Catalog Number: DYNJ910956; 28) TOURO OR LAMINECTOMY PACK, Catalog Number: DYNJ912687
FDA Recall Z-3269-2026 · Posted 2026-09-24
Advertisement
Recalling FirmMedline Industries, LP
Recall StatusOpen, Classified
Product CodeOJH
Date Initiated2026-08-10
Product Quantity4013 kits
Reason for Recall
The kits contain neurological sponges with the potential for unexpected variability of endotoxin levels. Potentially elevated endotoxin levels may present health risks that could require medical intervention, such as neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.
Firm Action
An "URGENT MEDICAL DEVICE RECALL" notification letter dated 8/10/26 was sent to customers.
REQUIRED ACTION:
1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product.
2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form.
Website link: https://recalls.medline.com Recall Reference #: R-26-159-FGX1
Recall Code:
3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable.
4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product.
If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Root Cause
Nonconforming Material/Component
Distribution Pattern
US Nationwide.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.