Fisher Scientific Sure-Vue Urine hCG, Model/ Part: FHC-102/23900526; Henry Schein, Inc. One Step + Urine Cassette Test, Model/ Part: FHC-102/9004074, FHC-102/9004073; Alere hCG Cassette (25mIU/mL) in Urine, Model/ Part: FHC-102/92217; Cardinal Health hCG Cassette Rapid test CLIA Waived, Model/ Part: FHC-102/B1077-22; Medline HCG Pregnancy Test Cassette, Model/ Part: FHC-102/MPH12025HCG; Pro Advantage By NDC Urine Pregnancy Cassette Device, Model/ Part: FHC-102/P080063;

FDA Recall Z-3274-2026 · Posted 2026-09-25

Recalling FirmAlere San Diego, Inc.
Recall StatusOpen, Classified
Product CodeJHI
Date Initiated2026-08-20
Product Quantity18,139,960

Reason for Recall

hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.

Firm Action

On 8/20/2026, correction notices started being emailed to customers and distributors informing them to take the following actions: 1) In situations where an unexpected test result is obtained, consider all limitations, including peri-menopause that may result in low levels of hCG. A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated, and an alternate confirmatory test that supports a timely clinical decision has been considered. 2) If you have forwarded affected products to other laboratories within your distribution network, notify them of this Field Corrective Action and provide them with a copy of this letter. 3) Retain this letter for your records. 4) Acknowledge receipt of this letter. Your organization s reply provides us with the necessary documentation to monitor the effectiveness of the field action notification. If you have questions regarding this information, U.S. Customers please contact the firm's Technical Services at 877-308-8289

Root Cause

Device Design

Distribution Pattern

Worldwide - US Nationwide distribution including in the states of IL, PA, FL, GA, VA, NY, TN, AZ, ID, ND, WI, AL, LA, NJ,MT, NE, OH, PR, IN, MI, WY, MA, CA, TX, HI, KS, KY, AR, CO, NC, WA, IA, MN, OK, OR, NM, SC, GU, MD, NV, MS, WV, SD, VT, DE and the countries of United Kingdom, Canada, Italy, Ireland, Spain, Portugal, New Zealand, Hong Kong, Australia, Papua New Guinea, Fiji, Guam, Sweden.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.