Terminated

Brilliance 64 Computed Tomography System, Model #78231, Serial Number: 9592. The Brilliance 64 is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at difference angles and planes. This device may include signal analysis and display equipment, patient, and equipment supports, components and accessories.

FDA Recall Z-3281-2011 · Posted 2011-09-27

Recall StatusTerminated
Product CodeJAK
Date Initiated2011-06-02
Date Terminated2013-03-15
Product Quantity1 unit

Reason for Recall

Component defect. Philips Medical decided to conduct a recall field /correction (rework) of the mounting key locking threaded insert of the Detector Module system (DMS) bolt on one (1) Brilliance 64 CT X-ray unit. The firm initiated this action following Philip's receipt of the customer's complaint reporting that the retaining inserts were pulling out of the rotor during the removal of the UDMS

Firm Action

Philips sent a Field Change Order (FCO) 72800519 on June 13, 2011 to the affected customer. A Field Service Engineer visited the customer site and installed a hardware upgrade to the affected Brilliance 64 device system (Serial Number: 9592). The correction action was completed on August 9, 2011. For questions regarding this recall call 1-440-483-7000.

Root Cause

Other

Distribution Pattern

Nationwide Distribution including PA

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →