Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Clinical Trial

Effects of Mechanical Insufflation-Exsufflation With Optimized Settings on Wet Mucus Volume During Invasive Ventilation

NCT: NCT06491017 · RECRUITING

NCT IDNCT06491017
StatusRECRUITING
Start Date2026-05-08

Brief Summary

Retention of airway secretions is a frequent complication in critically ill patients requiring invasive mechanical ventilation (MV).This complication is often due to excessive secretion production and ineffective secretion clearance. Mechanical insufflator-exsufflator (MI-E) is a respiratory physiotherapy technique that aims to assist or simulate a normal cough by using an electro-mechanical dedicated device. A positive airway pressure is delivered to the airways, in order to hyperinflate the lungs, followed by a rapid change to negative pressure that promotes a rapid exhalation and enhances peak expiratory flows. However, there is no consensus on the best MI-E settings to facilitate secretion clearance in these patients. Inspiratory and expiratory pressures of ±40 cmH2O and inspiratory-expiratory time of 3 and 2 seconds, respectively, are often used as a standard for MI-E programming in the daily routine practice, but recent laboratory studies have shown significant benefits when MI-E setting is optimized to promote an expiratory flow bias. The investigators designed this study to compare the effects of MI-E with an optimized setting versus a standard setting on the wet volume of suctioned sputum in intubated critically ill patients on invasive MV for more than 48 hours.

Official Source

View on ClinicalTrials.gov →

Data sourced from ClinicalTrials.gov API. For the most current status, refer to the official record.

↑