Noninvasive Blood Pressure Validation Study
NCT: NCT07774169 · COMPLETED
Brief Summary
The purpose of this study is to test how accurately the Perin Health Patch, a small wearable sensor worn on the chest, measures blood pressure without using an arm cuff. Adults 22 to 90 years of age wore the patch while trained observers took blood pressure readings at the same time using a standard arm cuff and stethoscope. The patch readings were then compared to the standard readings to see how closely they match, following recognized international testing standards for blood pressure devices (ISO 81060-2 and IEEE 1708). Participants visited one of three study sites up to three times over about one week. At the final visit, participants also did light resistance band exercise to briefly raise their blood pressure and relaxation activities to lower it, so the patch could be tested while blood pressure was changing.
Frequently Asked Questions
What is Noninvasive Blood Pressure Validation Study?
Noninvasive Blood Pressure Validation Study is a clinical trial registered under NCT07774169. Current status: COMPLETED.
What is the status of NCT07774169?
The current status of NCT07774169 (Noninvasive Blood Pressure Validation Study) is: COMPLETED.
When did Noninvasive Blood Pressure Validation Study start?
Noninvasive Blood Pressure Validation Study started on 2026-01-15.
Related Devices
Official Source
View on ClinicalTrials.gov →Data sourced from ClinicalTrials.gov API. For the most current status, refer to the official record.