Guider Softip XF; GUIDER MPXF 5F 90CM, Model number: M003100630. Cardiology: Boston Scientific Guider Softip XF Guide Catheters are intended to facilitate the placement of interventional devices into the neurovascular system.

FDA Recall Z-0344-2017 · Posted 2016-11-02

Recalling FirmStryker Neurovascular
Recall StatusTerminated
Product CodeDQY
Date Initiated2016-10-05
Date Terminated2017-03-23
Product Quantity1478 devices, all models and lot numbers.

Reason for Recall

Firm became aware of potentially defective Guider Guide Catheters due to a manufacturing non-conformance that could lead to hub leaks in the resulting product.

Firm Action

Stryker Neurovascular sent an Urgent Medical Device Voluntary Recall letter dated October 5, 2016, to all affected customers by Federal Express. Letters advised of the reason for recall and models and devices affected and requested that all affected devices in inventory be returned to Stryker. Customers were advised to circulate the Field Safety Notice internally to all interested/affected parties and to complete the attached customer response form and return the form to their nominated Stryker Representative or to [email protected]. For questions regarding this recall call 510-413-2106.

Root Cause

Employee error

Distribution Pattern

Worldwide Distribution - US including AK, IL, OH and Internationally to: ARGENTINA AUSTRALIA AUSTRIA BELGIUM BRAZIL CANADA CHILE CZECH REPUBLIC ECUADOR EGYPT FRANCE GERMANY GREECE HUNGARY INDIA IRAN ISRAEL ITALIA JAPAN MEXICO NETHERLANDS NORWAY PERU PHILIPPINES POLAND RUSSIA SERBIA SINGAPORE SLOVAKIA SOUTH AFRICA SOUTH KOREA SPAIN SWEDEN SWITZERLAND THAILAND TUNISIA TURKEY UK URUGUAY VIETNAM.