Terminated

Bausch & Lomb Millennium 20 Ga. High Speed Posterior Vitrectomy Pack, Catalog #CX4920, containing 1/20-gauge high speed vitrectomy cutter as well as other accessories. Catalog #USSCX4920A is also included in the recall, which is used by customers for ordering only . There is no labeling associated with this catalog number. This catalog number is used for ordering Cat. #4920 along with Cat. #MVS1011C-20g, Light Pipe (which is not under recall). These two products are packaged separately and are not bundled in any way. They are merely shipped at the same time. The firm name on the label is Bausch & Lomb Incorporated, Rochester, NY.

FDA Recall Z-0496-2007 · Posted 2007-05-24

Recalling FirmBausch & Lomb Inc
Recall StatusTerminated
Product CodeHQE
Date Initiated2006-12-14
Date Terminated2009-04-07
Product QuantityCat. CX4920 - 4,475 packs; and Cat. CX4920A - 40 packs

Reason for Recall

The cutter tip could break during a procedure

Firm Action

Recall letters were issued via e-mail to surgeons on 12/14/06 and via overnight mail on 12/14/06 to surgeons and operating room supervisors requesting immediate segregation and return of the product.

Root Cause

Other

Distribution Pattern

Worldwide distribution ---- including USA and countries of Canada, China, Japan, Australia, Hong Kong, India, Malaysia, New Zealand, Singapore, Korea, Taiwan, Thailand, Argentina, Brazil, Mexico, and The Netherlands.