Terminated

SKBM Microkeratome cutting tool

FDA Recall Z-0612-03 · Posted 2003-03-06

Recall StatusTerminated
Product CodeHNO
Date Initiated2002-11-18
Date Terminated2003-07-08
Product Quantity567

Reason for Recall

Degradation of glue causes glass movement resulting in misalignment that can cause deeper cuts than anticipated.

Firm Action

Customers have been notified in writing of this recall action (11/18/2002), and instructed to return all SKBM and/or Lasitome microkeratome products. All returned product will be physically segregated and dispositioned by Alcon. Recall Effectiveness checks will be attempted between 10%-100% of the consignees.

Root Cause

Other

Distribution Pattern

Distribution was nationwide and to the following foreign countries: Spain, Taiwan, UK, Mexico, Brazil, India, France, Ireland, China, Lebanon, Germany, Australia, Italy, Saudi Arabia, Greece, Japan, Netherland, Turkey, Korea, Romania, Latvia, Belgium, Switzerland, Malaysia, Thailand, Portugal, Phillipines, Canada, Sweden, Hungary, Argentina, Poland, Norway, Peru, Russia, Venezuela, Vietnam, Ecuador, Ukraine, Cyprus, Singapore, Chile, Nepal, Honduras, Colombia, and the Czech Republic.