Terminated

Bausch & Lomb Stellaris PC w/ Laser Upgrade BL1433 (Stellaris PC w v3 Software) with BL2272 (Laser Upgrade kit) Stellaris PC Vision Enhancement System with integrated laser module. System is used for the surgical treatment/removal of cataractous lens, vitreous removal and certain treatments of the posterior portion of the eye.

FDA Recall Z-0863-2015 · Posted 2014-12-23

Recalling FirmBausch & Lomb Inc
Recall StatusTerminated
Product CodeHQF
Date Initiated2014-11-04
Date Terminated2017-03-23
Product Quantity79

Reason for Recall

A software anomaly was identified where the system may automatically transition from "Standby" to "Ready to Fire" state while in surgical use.

Firm Action

Bausch & Lomb contacted customers by phone on November 4, 2014. Customers were informed of the conditions required for the device to malfunction. Customers were also notified that a software update to address the situation will be installed in their system within the next few weeks, and that the firm's service team will be scheduling a visit to each facility to install this software update. Customers were also told that in the meantime they may continue to use the device. For further questions please call (585) 338-6839.

Root Cause

Software Manufacturing/Software Deployment

Distribution Pattern

Worldwide Distribution - USA Distribution including to the states of AL, CA, CT, FL, GA, HI, IL, IN, KS, MI, MN, MO, MS, NH, NY, OH, OK, PA, TX, WI and WV., and to the countries of Austria, Australia, Belgium, China, Czech Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, India, Ireland, Italy, Jordan, Malaysia, Morocco, Netherlands, Norway, Portugal, Singapore, Spain, Sweden, Switzerland, Turkey and the United Arab Emirates.