Digital Medical X-ray Imaging System, Model: uDR 596i, uDR 592h
FDA Recall Z-1075-2022 · Posted None
Advertisement
Recalling FirmShanghai United Imaging Healthcare Co., Ltd.
Recall StatusTerminated
Product CodeKPR
Date Initiated2019-07-30
Date Terminated2022-05-17
Product Quantity4
Reason for Recall
X-ray imaging system positioning image and protocol label is reversed for Flexion and Extension on C-Spine and L-Spine, this will cause the image to be incorrectly labeled and may cause the patient to have a repeat exposure.
Firm Action
On 07/30/19, Customer Notification Letters started being distributed to customers. Customers were asked to contact the firm if they found an inconsistency between the protocol name and the positioning image. The following phone numbers were provided: (855) 221-1552 (toll free) or (832) 699-2799 (direct).
The firm issued Field Correction Instructions to Service Engineers to modify the protocol name of Flexion and Extension on C-Spine & L-Spine with the corrected guidelines. In addition, the firm committed to permanently fix impacted systems with a software update.
Root Cause
Software change control
Distribution Pattern
US Nationwide distribution in the state of TX.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.