Open, Classified

Trevo Trak 21 Microcatheter, REF: 90338. TREVO NXT 3X32+TRAK21 2-PACK - CE, REF: 91412; TREVO NXT 4X28+TRAK21 2-PACK - CE, REF: 91413; TREVO NXT 4X41+TRAK21 2-PACK - CE, REF: 91414; TREVO NXT 6X37+TRAK21 2-PACK - CE, REF: 91415; TREVO NXT 4X28+TRAK21+CAT6 3-PACK - CE, REF: SFP914170 TREVO NXT 6X37+TRAK21+CAT6 3-PACK - CE, REF: SFP914190. Used in the placement of devices and/or fluids, such as contrast media, into the peripheral, coronary, and neuro vasculature during diagnostic and/or therapeutic procedures.

FDA Recall Z-1386-2022 · Posted None

Recalling FirmStryker Neurovascular
Recall StatusOpen, Classified
Product CodeDQO
Date Initiated2022-06-03
Product Quantity71

Reason for Recall

Microcatheter product label on the carton is missing the microcatheter product drawing/illustration.

Firm Action

On 06/03/22 Field Action notifications were mailed to customers who were informed of the following: 1) The affected items may continue to be used. Refer to the inner pouch label for the product drawing/illustration. 2) Circulate and maintain awareness of this Customer Notification internally to all interested/affected parties. 3) Inform the recalling firm if any of the subject devices have been distributed to other organizations. If yes, provide contact details so that the recalling firm can inform the recipients appropriately. 7. Please inform the recalling firm of any adverse events concerning the use of the subject devices. 8. Complete and return the the customer response form. The response form or questions can be addressed to your local firm representative or via email to [email protected]

Root Cause

Labeling mix-ups

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of CA, OR, PA, NY, DC, IL, WA, NH, NJ, OK, FL, AZ, AL, GA, TX, WI, VA and the countries of DE, IT, AT.