uCT 530 Computed Tomography X-Ray System
FDA Recall Z-2472-2020 · Posted None
Advertisement
Recalling FirmShanghai United Imaging Healthcare Co., Ltd.
Recall StatusOpen, Classified
Product CodeJAK
Date Initiated2020-05-08
Product Quantity4 units
Reason for Recall
Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.
Firm Action
The firm sent letters to their customers by email on 5-22-2020. The letter informs customers of the two potential safety issues with the device, and notes that these problems may cause image artifacts and scanning workflow interruptions. Risks to patient health are noted to include the possible need to rescan the patient resulting in an additional radiation dose and the possible need for additional contrast medium. Customers were informed that United Imaging has released a field update for the uCT 530 to address the issues.
If customers observe image artifacts, they are requested to reboot the system and continue the reconstruction. If the scout scanning is interrupted, customers are instructed to try repeating the scout scanning. If problems persist, customers should contact their United Imaging service support team or call the United Imaging Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct).
Root Cause
Software design
Distribution Pattern
Distributed to Texas.