Computed Tomography X-ray System, Model: uCT 550, GTIN: 06971576831036 Intended use The uCT Computed Tomography X-ray System uCT 530/uCT 550 is a computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck and vascular).
FDA Recall Z-2641-2020 · Posted 2020-07-20
Advertisement
Recalling FirmShanghai United Imaging Healthcare Co., Ltd.
Recall StatusOpen, Classified
Product CodeJAK
Date Initiated2020-05-23
Product Quantity3
Reason for Recall
Two issues were identified with the computed tomography x-ray system including a service function which may cause false marking of a bad channel resulting in ring artifacts, and potential intermittent scout scanning interruption due to occasional angle signal drift.. If these problems occur, it may be necessary to rescan the patient resulting in an additional dose of radiation and the possible need for additional contrast medium.
Firm Action
On 05/23/2020, Customer Notification Letters dated 5/22/2020 were emailed to customers. Customers were asked to do the following:
For safety issue 1:
If you observe image artifacts, please reboot the system and continue the reconstruction. If the image artifacts persist, please contact the recalling firm's service support team or call the Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct).
For safety issue 2:
If the scout scanning is interrupted, try repeating the scout scanning. If problem persists, please contact the recalling firm's service support team or call the Customer Contact Center at (855) 221-1552 (toll free) or (832) 699-2799 (direct).
Customers were asked to notify all users of the affected products via Customer Notification Letter.
On 8/30/2024, a correction notice was sent identifying an additional serial number.
Root Cause
Software design
Distribution Pattern
US Distribution to state of: TX
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement