Bausch & Lomb Stellaris Optimized Stability Vacuum Pack, Ref: BL5112, packaged 6 units/case, RX, Sterile. Bausch & Lomb Incorporated, Rochester, NY For use with the Bausch & Lomb Stellaris Vision Enhancement System for the phacoemulsification of an opacified crystalline lens during anterior segment surgery
FDA Recall Z-3089-2011 · Posted 2011-08-30
Advertisement
Recalling FirmBausch & Lomb Inc
Recall StatusTerminated
Product CodeHQC
Date Initiated2011-07-27
Date Terminated2013-06-21
Product Quantity2/6-unit cases
Reason for Recall
The aspiration channel on the affected vacuum cassettes may be blocked due to a molding malfunction
Firm Action
Bausch & Lomb Inc. notified all affected customers by direct telephone contact on July 27, 2011 and a follow-up FAX or e-mail dated July 27, 2011
The letter included; affected product, problem, and requests customers return the product. An acknowledgement form was included to complete and return with the product.
For question on this recall contact Bausch & Lomb at (949) 521-7895.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Nationwide Distribution.