Bausch + Lomb Stellaris Elite 23 GA Vit Cutter, Model BL5627, 7500 cpm, packaged in sterile pouches, 6/shipper.

FDA Recall Z-3268-2018 · Posted None

Recalling FirmBausch & Lomb Inc
Recall StatusTerminated
Product CodeHQE
Date Initiated2018-08-08
Date Terminated2019-08-28
Product Quantity103.33 cases

Reason for Recall

The back cap separates from the body of the vitrectomy cutter.

Firm Action

Recall letters flagged URGENT MEDICAL DEVICE VOLUNTARY RECALL dated 8/8/2018 were issued 8/8/2018 via certified mail return receipt to the U.S. consignees. Letters for foreign consignees were dated 8/7/2018 and were flagged URGENT FIELD SAFETY NOTICE.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Distribution was made to AL, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, MA, ME, MO, MT, NC, NJ, NY, OH, OK, OR, PA, TN, TX, UT, and WV. There was government distribution but no military distribution. Foreign distribution was made to Russia, Romania, Poland, The Netherlands, Jordan, Italy, Israel, Switzerland, Hungary, Greece, Guadeloupe, Great Britain, France, Spain, Singapore, Malaysia, and Thailand.